The CEC-20 is a new approach to CE

and CEC in one package, which

addresses the needs of both techniques.

 

 

       CEC-20 Capillary Elophoresis System                                                   for CE with 45 position autosampler and UV detector                          CEC-20+ Capillary ElectroChromatography CEC System                            with 45 position autosampler and UV detector

Options: Capillary Fluorometer detector ✻✻

 

特色

1. It has a 45-position autosampler, capillary mounting, UV-VIS detector and electrode mounting hardware.

2. It incorporates an on-board computer and software for Data Acquisition and Control as well as Data Storage.

3. It features temperature control of the sample chamber, carousel and column for reproducible migration times and temperature-sensitive biological samples.

4. All sample contact materials other than the Pt electrodes are bio-compatible plastics.

5. Flexible Configurations: 

The CEC-20 can be configured with a wide range of column lengths(columns may be as short as 15 cm to greater than 2 meters in length). It has a unique patent-pending capillary mounting system for the on-broad absorbance detector.

The CEC-20 addresses the requirements for open capillaries and for CEC packed or continuous bed columns requiring either low pressure or high pressure by offering two types of pumps for column filling and conditioning.

6. Reliable, rugged, and accessible low maintenance design

The CEC-20 is made from strong lightweight materials that provide a compact yet rugged design with a small footprint. All reagents and buffers are enclosed within an easily accessible work area with short paths from the autosampler to column to detector. The CEC-20 is a narrow profile system that incorporates all of the necessary reagents and buffers for CE/CEC column preparation, derivatization and cleaning that are part of the CE/CEC process.

 

規格

Sampling Module

Automated 45 position sample tray suitable for 12 nm vials(2 mL size)

Sample Contact material

All biocompatiable materials are used in the sample contact areas

Automated Sampling valves

Automated sampling Injection ports along with multiple port buffer, waste, cleaning, derivatization stations are provided as a standard feature and are plumbed for standard injections

Sampling Modes

Both Timed Pressure injections and electrokinetic injections are supported

Pumps

Standard pumps will deliver up to 8 bar; optional HPLC pumps will deliver up to 400 bar(CEC-20+)

Detection Module

Proprietary on column UV/VIS capillary absorption monitor(190 nm to 940 nm)

High Voltage Power Supply

Software-controlled +/- 32KV power supply with current clamping safety feature

Temperature Control

User selected operational temperature control of capillary and sample area(Peltier Cooler/Heater)

Data Collection

Integrated Computer with embedded JMST HydrogenTM Data Acquisition Control and Reduction Software

Input/Output connections

Standard Computer interface is provided through the integral Computer. Ethernet TCP/IP protocol is supported for connection to networks and other computers for remote controls and data collection

Displays

Two 5 x 4 LED displays, 8digits each and 8 keys each. An external Computer Monitor is required for connection to the system(1280 x 1024 LCD monitor is recommended but not included with the system, except as a special order).

Size

63 cm (height) x 32.5 cm (width) x 61 cm (Depth)Approx. 45kg (100 lb. shipping weight)

 

 

 


 

 

 

 

<<回高效能液相層析儀(HPLC)>>   <<回HPLC分析儀器與層析軟體>>