Characteristics of Rheodyne Manual Sample Injectors

Model1

Scale

Partial Filling Volumes (Range)

Sample Loop

Sizes (Range)

Main material

Liquid-
contact

Materials

Max.

MPa2

Max.

T ℃

MBB3

Type and Capabilities

切換示意圖

7725, 7725i

Analytical

1uL ~ 2.5 mL

2uL ~ 5.0 mL

316 SST

316 SST, Vespel,

ceramic, PEEK

48

80o

Yes

Dual Mode

Can load the loop by two methods:

(1) Partial filling syringe determines volume without wasting sample

(2) Complete filling loop determines volume by overfilling loop

A

7125

5uL ~ 5.0 mL

48

80o

No

7125-081

Not Applicable

10uL ~ 5.0 mL

Titanium

Titanium, Tefzel, ceramic, PEEK

34

50o

9725, 9725i

1uL ~ 5.0 mL

2uL ~ 10 mL

PEEK

PEEK, Tefzel, ceramic

34

50o

Yes

8125

(Low dispersion)

Micro

0.1uL ~ 500uL

5uL ~ 1.0 mL

316 SST

316 SST, Vespel

ceramic, PEEK

48

80o

No

3725,3725i

Preparative

100uL ~ 10mL

2.0 mL  ~ 20 mL

PEEK

PEEK

28

50o

Yes

3725(i)-038

316 SST

316 SST, PEEK

34

50o

Yes

7010

Analytical

Not Applicable

5uL ~ 5.0 mL

316 SST

316 SST, Vespel

48

150o

No

Single Mode

Can load the loop by one method:Complete filling loop determines volume by overfilling loop

B

7010-087

10uL ~ 5 mL

Titanium

Titanium, Tefzel

34

80o

No

9010

5uL ~ 10 mL

PEEK

PEEK, Tefzel, ceramic

34

50o

No

7410

(Internal Loop)

(Low dispersion)

Micro

0.5uL ~ 5uL

316 SST

316 SST, Vespel

48

150o

No

C

7520

(Internal chamber)

0.2uL ~ 1uL

316 SST

316 SST, Vespel

48

80o

No

D